| Measurement range |
1: 0 – 5 mg/L
2: 0 – 5 mg/L (Including pH compensation) |
| Accuracy |
0.01 ppm (0 – 0.5 ppm) ; 0.02 ppm (0.5 – 2 ppm) ; 0.1 ppm (2 – 5 ppm) |
| Resolution |
0.01 mg/L |
| Repeatability |
± 1% FS |
| Temperature compensation |
0 – 85 °C |
| Automatic pH compensation |
5 – 9 pH (Type 2) |
| First setup time |
4 hours |
| Process flow rate |
Min. 0.8 l/min |
| Auto cleaning |
Rotating glass ball |
| Electrode material |
Platinum (Cathode), Copper (Anode) |
| Operating pressure |
0 – 10 Kg f/cm2 |
| Operating temperature |
0 – 85 °C |
| Response time |
3 secs |
| Protection |
IP 68 |
| Temperature Sensor |
Pt 100 |
| Temperature measurement range |
0 – 65 °C |
| Temperature accuracy |
± 0.5 °C ; ± 0.1 °C (Optional) |
| Housing |
Acrylic |
| pH Sensor |
|
| Measurement range |
0 – 14 pH |
| Accuracy |
± 0.05 pH |
| Resolution |
0.01 pH |
| Repeatability |
± 1% FS |
| Operating pressure |
Max. 10 Kgf/cm2 (Optional) |
| Operating temperature |
0 – 85 °C |
| Process flow rate |
0.1l /min |
| Response time |
1 sec |
| Dimension |
1: 152(W) x 168(H) x 118(D) mm
2: 215(W) x 300(H) x 155(D) mm |
| Weight |
1 analyzer: approc. 300g
2 analyzer: approc. 500g ; cable: 80 g/m |